Abstract

Two rapid and popular methods—capillary electrophoresis (CE) and high-performance liquid chromatography (HPLC) have been compared for analysis of cotinine in human urine. Cotinine was analyzed in less than 7 min, with detection limits of 5 and 3.2 ng mL−1 for CE and HPLC, respectively. The performance of the methods was evaluated in terms of sensitivity, specificity, precision, accuracy, and limits of detection and quantification. Calibration plots were linear in the range 50–4,000 ng mL−1, at least, and mean recoveries were satisfactory for both techniques. The methods were successfully used for quantification of cotinine in urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.