Abstract
To compare intraocular pressure, anterior segment optical coherence tomography, and ultrasound biomicroscopy parameters over 3 months after panretinal photocoagulation (PRP) for proliferative diabetic retinopathy after 1 of 2 sittings by conventional laser (half PRP) and a single sitting of Pattern Scan Laser (PASCAL) PRP. This was a prospective, randomized, interventional study. All tests were performed at baseline, and at 1, 6, and 24 hours, and 1, 4, 8, and 12 weeks after PRP. The intraocular pressure at 1 hour and 6 hours after PRP was significantly raised in both groups. Mean intraocular pressure was 21.17 ± 4.01 mmHg after PASCAL and 17.48 ± 3 mmHg after conventional laser at 1 hour, P < 0.001. On anterior segment optical coherence tomography, conventional laser PRP caused a more significant narrowing of angle-opening distance (AOD750) and trabecular-iris space area (TISA 500), P = 0.03 and 0.04, respectively, on Day 1. Ultrasound biomicroscopy showed a significantly narrow angle in both groups on Day 1. A significant increase in ciliary body thickness was observed in both groups, with 57.1% of PASCAL and 100% of conventionally treated eyes showing ciliary effusion on Day 1 that decreased but persisted for the next 3 months. Performing PRP in sittings, prescribing previous glaucoma medications in patients at risk, and recording intraocular pressure an hour after the PRP could decrease complications.