Abstract

The purpose of this study was to evaluate the effect of denture cleansers on the surface roughness, hardness and color stability of two polyamides (Valpast, Deflex), a butadiene styrene copolymer PMMA (Rodex), and PMMA polymer as a control group (Paladent). Each material was divided into 5 sub-groups (n=7) as two control and three test groups. Three test groups were immersed for 20 days in commercially available three denture cleansers (CO-Corega, PR-Protefix, VA-Valclean). Two-way analysis of variance and Tukey's post hoc HSD test were used to evaluate surface roughness and hardness data (α=0.05). ΔE, ΔL*, Δa*and Δb* mean values were used for ANOVA, Tamhane test was used as post hoc. Polyamides showed low hardness and high roughness before and after immersion. A significant decrease in hardness was observed for all resins except Rodex after immersion (p<0.05). The denture cleansers changed the roughness, hardness and color of some resins.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.