A compact photomultiplier housing is described. The housing can provide controlled cooling to the photomultiplier tube down to −90°C. The cooling is accomplished by flowing liquid nitrogen cooled helium gas through a series of coils which envelope the photocathode portion of the tube. The temperature is controlled by controlling the flow of the gas with a fine adjustable needle valve. The temperature is measured near the photocathode of the photomultiplier by a calibrated thermistor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.