Abstract
A fiber-optic Fabry–Perot (FP) based relative humidity (RH) sensor was proposed and experimentally investigated by using Poly(N-isopropylacrylamide) (PNIPAM) hydrogel to connect two ends of single-mode optical fiber (SMF) to formulate a PNIPAM FP cavity. The water vapor will be absorbed by the PNIPAM hydrogel, which will alter both the refractive index (RI) and dimension of the PNIPAM FP cavity, resulting in the change of output resonance intensity in the reflective spectrum. The measure RH sensitivity is 1.634 nm/%RH within RH range from 45 to 75 % with good linearity of 0.9897 and excellent repeatability. The influence of PNIPAM hydrogel concentration on the RH sensitivity has been investigated - the higher the concentration of PNIPAM hydrogel, the higher the RH sensitivity of the sensor. The developed sensor can effectively monitor human respiratory with response and recovery time of 2.29 s and 1.90 s, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.