Acoustic-resolution photoacoustic imaging enables unique high-resolution visualization of chromophores in deep tissue. However, most current systems are bulky and use complex optical components for illumination. In this study, we developed a compact acoustic-resolution photoacoustic imaging system with simple fiber-based illumination. Around the ultrasound sensor attached to an acoustic lens, four optical fibers were placed in four directions 90° apart, and the sensor plane was covered with an optically and acoustically transparent plastic spacer. The imaging head is compact (diameter, 32 mm; height, 27 mm) and lightweight (32 g) and is alignment-free, thus being suitable for in vivo applications. We assessed the uniformity and symmetry of photoacoustic excitation both by Monte Carlo simulation and experiments. A phantom study showed lateral and axial resolutions of the system to be 148 and 66 µm, respectively. With the system, high-contrast visualization of blood vessels in rat skin was demonstrated in vivo.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.