Abstract
In an improved Andersson-Braun (A-B) rem meter, Pb(n, xn) reactions at high neutron energies increase rem meter sensitivity when lead replaces a small part of the neutron-moderating material, polyethylene. Monte Carlo calculations were carried out using the Los Alamos High-Energy Transport code (LAHET) and Monte Carlo Neutron Transport (MCNP) code for both standard and lead-modified A-B rem meters with monoenergetic neutrons (up to 2.1 GeV). Experiments for medium and high energy tests (23.6 MeV-2.1 GeV at nine energy values) verified the results of theoretical calculations of response enhancement. Thus, Monte Carlo calculations that use the LAHET and MCNP codes prove to be powerful tools to simulate new neutron detectors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.