Abstract

A new sample preparation method for MALDI based on the use of a mixture of the two commonly used matrices, DHB and CHCA, is described. The matrix mixture preparation results in increased sequence coverage and spot-to-spot reproducibility for peptide mass mapping compared to the use of the single matrix components. This results in more reliable protein identification in proteomics studies and facilitates automated data acquisition. This method shows better tolerance towards salts and impurities, eliminating the need for pre-purification of the samples. It has also been found to be advantageous for the analysis of intact proteins, and especially for glycoproteins. The mixture allows the presence of rather high concentrations of urea in the sample solutions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.