Abstract
A novel methodology based on column switching LC–MS, using restricted access material (RAM), for an integrated sample clean-up and analysis of endocrine disrupting compounds in sediment samples is described. The use of RAM precolumns, that combines size exclusion and reversed-phase retention mechanisms, enables fast on-line clean-up of sediment extracts and sensitive determination of alkylphenolic compounds, bisphenol A and steroid sex hormones at low ppb level (LODs=0.5–5 ng/g). Different LiChrospher ADS RAM precolumns (Merck, Germany) with C 4, C 8 and C 18, respectively, modification of inner pore surface were tested. ADS C 4 precolumns gave the best results in terms of recovery, selectivity and sensitivity, eliminating efficiently matrix components and consequently reducing ion suppression effects. Except for the most polar compounds, all compounds exhibit complete recovery with RSD from 0.87 to 15%. A complete analysis, including efficient pressurized liquid extraction (PLE) of target compounds, on-line clean-up, chromatographic separation and MS detection takes approximately 2 h, which is a significant improvement in comparison to the methods reported previously.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.