Abstract

Restricted access materials (RAM) based on alkyl diol silica (ADS) porous particles were evaluated for the sample clean-up in tetracyclines analysis in milk and water samples by liquid chromatography. RAM with C 4, C 8 and C 18 bonded to the inner pore surfaces were tested, with C 8 providing the best performance. A switching column LC system which combines an ADS C 8 RAM column and an analytical C 18 column connected to a fluorimetric detector was used for tetracyclines analysis. The RAM clean-up removed large peaks that otherwise appeared in the initial time window of the chromatograms, attributed to proteins in milk samples and humic substances in water samples. Thus, quantification of analytes in real samples, especially of the most polar compounds such as oxytetracycline and tetracycline, was clearly improved.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.