Abstract

A rapid immunochromatographic assay was developed to detect ricin. The assay was based on the sandwich format using monoclonal antibodies (Mabs) of two distinct specificities. One anti-ricin B chain Mab (1G7) was immobilized to a defined detection zone on a porous nitrocellulose membrane, while the other anti-ricin A chain Mab (5E11) was conjugated to colloidal gold particles which served as a detection reagent. The ricin-containing sample was added to the membrane and allowed to react with Mab (5E11)-coated particles. The mixture was then passed along the porous membrane by capillary action past the Mab (1G7) in the detection zone, which will bind the particles that had ricin bound to their surface, giving a red color within this detection zone with an intensity proportional to ricin concentration. In the absence of ricin, no immunogold was bound to the solid-phase antibody. With this method, 50 ng/ml of ricin was detected in less than 10 min. The assay sensitivity can be increased by silver enhancement to100 pg/ml.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.