Abstract

In the last 10 years the atomic force microscope (AFM) has routinely been applied to the direct measurement of surface forces. A particle is often attached to the tip of the force-sensing cantilever in order to control the chemistry and the geometry of the interaction. Usually a spherical particle (colloid probe) is employed, and the measured interaction force is normalized by the radius, enabling comparisons to be made with both theory and other techniques. Here we report a simple method for characterizing the radius, surface roughness, and regularity of colloid probe particles. The colloid probe is “reverse imaged” by employing a surface with very sharp features. A standard calibration grating is suitable. No further equipment is required. The method is straightforward and will be of value to researchers employing the AFM for force measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.