Abstract

A previously published gas chromatography-mass spectrometry (GC-MS) method for the analysis of 7-aminoflunitrazepam (7-amino-FM2) in urine was subjected to an inter-laboratory collaborative study. The method was rigorously defined in terms of performance requirements, yet allowed some degrees of flexibility to individual analysts. The collaborative study followed the guidelines provided by the Association of Official Analytical Chemists International (AOAC International). Each of nine drug-abuse urine testing laboratories in Taiwan participated in the study to analyze 3 samples in duplicate. The repeatability relative standard deviation (RSDr) and the reproducibility relative standard deviation (RSDR) of the measurements of 7-amino-FM2 in urine were determined to be less than 9.91% and 14.84%, respectively. The method showed acceptable reproducibility, as evidenced by HORRAT values of 0.6 to 0.7. The method was thus proposed to be used by the testing laboratories in Taiwan.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.