Abstract

We propose a novel use of a non-destructive technique to quantitatively assess hydrogen concentration in zirconium alloys. The technique, called Cold Neutron Prompt Gamma Activation Analysis (CNPGAA), is based on measuring prompt gamma rays following the absorption of cold neutrons, and comparing the rate of detection of characteristic hydrogen gamma rays to that of gamma rays from matrix atoms. Because the emission is prompt, this method has to be performed in close proximity to a neutron source such as the one at the National Institute of Technology (NIST) Center for Neutron Research. Determination shown here to be simple and accurate, matching the results given by usual destructive techniques such as Vacuum Hot Extraction (VHE), with a precision of ±2mgkg−1 (or wtppm). Very low levels of hydrogen (as low as 5mgkg−1 (wtppm)) can be detected. Also, it is demonstrated that CNPGAA can be applied sequentially on an individual corrosion coupon during autoclave testing, to measure a gradually increasing hydrogen concentration. Thus, this technique can replace destructive techniques performed on “sister” samples thereby reducing experimental uncertainties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.