Abstract

Hemoglobin A1c (HbA1c), also known as glycated hemoglobin, is a blood test used to evaluate and track a patient's blood sugar levels over the previous 2-3 months. We have compared the analytical performance of the D10 hemoglobin (HPLC) testing system to that of the immunoturbidimetric technique, which is a light-scattering immunoassay. To assess the clinical risk assessment between two methods (Compare the two Immunoturbidometric methods (AU680) vs. HPLC method (D10)) in hyperglycemic patients and assess the acceptability of the respective methods in the Clinical biochemistry laboratory. The charge of the globins in Hb was used as the basis for the HPLC method used to measure HbA1c. HPLC detects and quantifies even the tiniest Hb fractions and the full spectrum of Hb variants. HbA1c was measured using the immunoturbidimetric (AU 680 Beckmann coulter analyzer) and high-performance liquid chromatography (HPLC) techniques. Experiments also made use of immunoturbidimetric techniques (using an AU 680 Beckmann coulter analyzer equipment). There is no statistically significant difference in HbA1c readings between male and female patients, as measured by either the Immunoturbidimetric or HPLC techniques. The immunoturbidimetric and high-performance liquid chromatography techniques for estimating HbA1c yielded identical results. From the results of this study, we may deduce that both techniques are valid for estimating HbA1c. As a result, it may be suggested that both approaches can be used to estimate HbA1c in diabetic individuals.