Abstract

Background & aimsThe authors assess the diagnostic accuracy of the Transient Elastography-Controlled Attenuation Parameter (TE-CAP) in children of Southern China. Methods105 obese or overweight children and adolescents were enrolled in the diagnostic test of TE-CAP assessment of hepatic steatosis using MRI-PDFF. Hepatic steatosis grades S0-S3 were classified. Statistical correlation, agreement and consistency between methods were evaluated. The diagnostic efficiency of TE-CAP was evaluated. The authors used the cutoff value of TE-CAP to detect hepatic steatosis in another 356 children. ResultsThe Area Under Curve (AUC) of TE-CAP for grade ≥ S1, ≥ S2, and ≥ S3 steatosis were 0.975, 0.984, and 0.997, respectively. For detecting ≥ S1 steatosis, TE-CAP had a sensitivity of 96 % and a specificity of 97 %. For detecting ≥ S2 steatosis, TE-CAP had a sensitivity of 97 % and a specificity of 93 %. For detecting ≥ S3 steatosis, TE-CAP had a sensitivity of 1 and a specificity of 94 %. TE-CAP and MRI-PDFF had a linear correlation (r = 0. 0.87, p < 0.001). The hepatic steatosis was identified in 40.2 % (143/356) of children in which the obesity and overweight were 69.8 % (113/162) and 40.0 % (18/45). ConclusionTE-CAP showed excellent diagnostic accuracy in pediatric hepatic steatosis.