Abstract
Supraglottic devices have revolutionized the current practice of airway management. We compared the clinical performance of a recently introduced BlockBuster™ Laryngeal mask airway with i-gel® in adult patients under general anesthesia. Following Institutional ethical clearance, the present study was conducted on 62 patients belonging to American Society of Anesthesiologists physical status 1 and 2 of either sex in the age group of 20-60 years under general anesthesia. Patients were randomly assigned to i-gel® (I) and BlockBuster™ (B) groups (31 per group). Time for successful insertion, insertion success rate, ease of insertion, oropharyngeal leak pressures (OLPs), and complications were assessed. Mean insertion time of device was less in Group I (13.52 ± 2.58 s) than that of Group B (14.10 ± 2.04 s), which was neither clinically nor statistically significant (P = 0.330). OLP in Group B (24.52 ± 2.77 cm of H2O) was found to be significantly higher compared to Group I (20.81 ± 2.56 cm of H2O) with P < 0.001. Overall insertion and first attempt success was similar (i-gel® 31/31 [100%] and 29/31 [93.5%] and BlockBuster™ 31/31 [100%] and 29/31 [93.5%], respectively). Ease of insertion (P = 0.684) and complications (P = 0.782) of both the devices were comparable. Both the devices are useful and effective for airway management in adult under general anesthesia. Having a high OLP and comparable insertion time, risk of aspiration may be further reduced with the use of BlockBuster™ in comparison to i-gel®.