Abstract

A prototype Toxi.Prep (TP) system that utilizes solid-phase extraction (SPE) has been developed as a method for broad-spectrum drug screening and identification of tetrahydrocannabinol (THC) metabolites in urine. TP can simultaneously extract up to seven specimens while automating the process of sample extraction, washing, and elution onto a chromatogram. TP was compared with the Toxi. Lab A (TL-A) system for extraction of basic drugs only. In a blind study, 33 distinct drugs and metabolites were detected in 55 urines over 13 runs. Of the drug occurrences, 68.8% (141 of 205) were detected on both TP and TL-A. Of the 13 runs, quinidine and quinine, nortriptyline metabolites, and diphenhydramine were noted more frequently on TP than TL-A, whereas nicotine and metabolites, morphine, and methadone metabolites were more frequently noted on TL-A. Twenty specimens were analyzed for THC metabolites. Of the cases positive for THC metabolites, 100% (16 of 16) were positive by both methods. Time and motion studies for all runs proved an overall labor reduction for extraction and spotting by approximately 40%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.