Abstract

PurposeThe General Electric (GE) Swiftscan solution combines a new Low Energy High Resolution and Sensitivity collimator (LEHRS) with image processing (Clarity 2D) and tomographic step and shoot continuous mode. The aim of this study was to compare clinical and physical performances of this new technology in bone scintigraphy.MethodsPhysical phantom measurements were performed using GE LEHRS, GE Low Energy High Resolution (LEHR) and Siemens LEHR collimators. These measurements were associated with a prospective clinical study. Sixty-seven patients referred for bone scintigraphy were enrolled from February to July 2018. Each patient underwent two acquisitions consecutively on GE and Siemens gamma camera, using respectively Swiftscan solution and LEHR collimator.ResultsOn planar acquisitions, maximum sensitivity was 100 cts/MBq for Siemens LEHR. GE SwiftScan LEHRS and GE LEHR maximum sensitivity were respectively 9% and 22% lower. Using Clarity 2D, GE Swiftscan LEHRS spatial resolution was the best with 9.2 mm versus 10.1 mm and 10.6 mm for GE LEHR and Siemens LEHR collimators. In tomographic mode, the sensitivity of GE Swiftscan solution was superior to both LEHR systems (16% and 25% respectively for Siemens and GE). There was no significant difference in spatial resolution. In clinical use, signal was higher on Siemens system and noise was lower on GE Swiftscan solution. Contrast-to-noise ratios were not significantly different between the two systems. There was a significant image quality improvement with GE SwiftScan in planar images and in whole body scan. No significant difference in image quality was observed on SPECT images.ConclusionNew GE SwiftScan collimator design improved sensitivity compared to “classical” GE LEHR collimator without compromising resolution. GE SwiftScan solution enhances planar image quality with a better Clarity 2D resolution recovery and noise treatment. In SPECT mode, GE SwiftScan solution improves volumetric sensitivity without significant impact on image quality, and could lead to time or dose reduction.

Highlights

  • Most scintigraphic exams use Tc-99m radiolabeled pharmaceuticals and require Low-Energy High Resolution (LEHR) collimators

  • General Electric (GE) SwiftScan Low Energy High Resolution and Sensitivity collimators (LEHRS) and GE LEHR maximum sensitivity were respectively 9% and 22% lower

  • The sensitivity of GE Swiftscan solution was superior to both LEHR systems (16% and 25% respectively for Siemens and GE)

Read more

Summary

Introduction

Most scintigraphic exams use Tc-99m radiolabeled pharmaceuticals and require Low-Energy High Resolution (LEHR) collimators. General Electric Healthcare (GE Healthcare, Milwaukee, WI, USA) recently commercialized a SwiftScan solution for planar and SPECT acquisitions with a pair of new Low Energy High Resolution and Sensitivity collimators (LEHRS). SwiftScan solution combines LEHRS collimators and Clarity 2D image enhancement algorithm for resolution recovery and noise reduction in planar acquisitions. Clarity 2D is a 3 steps post-processing workflow on 2D images, automatically applied just after acquisition. This workflow is composed of a noise reduction with adaptive edge preservation, contrast enhancement and filtered image blending with native image. In SPECT mode, LEHRS collimators with step-and-shoot continuous mode enhance sensitivity, allowing data acquisitions during heads rotation.

Objectives
Methods
Results
Discussion
Conclusion