Abstract

[Purpose] The aim of this study was to investigate the clinical usefulness of a novel apparatus intended to improve the consistency and reliability of the deep tendon reflex (DTR) test. [Subjects] The participants included 50 healthy adults (25 male and 25 female) between 20 and 31 years of age who showed no knee abnormalities upon physical examination. [Methods] The intraclass correlation coefficient (ICC) was used to verify inter-rater reliability for two parameters examined in the DTR test using a newly designed apparatus. These parameters were the patellar reflex amplitude and first knee extension angle. Pearson’s product correlation coefficient was then used to examine the correlation between these two parameters. [Results] The inter-rater reliability analysis showed a high correlation between the examiners for both DTR parameters (ICCs = 0.91–0.96). In addition, a significant positive correlation was observed between the two parameters (r = 0.91). [Conclusion] The results show that it is possible to use the novel apparatus described herein to obtain reliable results in the DTR test.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.