Abstract
[Purpose] The aim of this study was to investigate the clinical usefulness of a novel apparatus intended to improve the consistency and reliability of the deep tendon reflex (DTR) test. [Subjects] The participants included 50 healthy adults (25 male and 25 female) between 20 and 31 years of age who showed no knee abnormalities upon physical examination. [Methods] The intraclass correlation coefficient (ICC) was used to verify inter-rater reliability for two parameters examined in the DTR test using a newly designed apparatus. These parameters were the patellar reflex amplitude and first knee extension angle. Pearson’s product correlation coefficient was then used to examine the correlation between these two parameters. [Results] The inter-rater reliability analysis showed a high correlation between the examiners for both DTR parameters (ICCs = 0.91–0.96). In addition, a significant positive correlation was observed between the two parameters (r = 0.91). [Conclusion] The results show that it is possible to use the novel apparatus described herein to obtain reliable results in the DTR test.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.