Abstract
To avoid malalignment of components during hip resurfacing arthroplasty (HRA), we used a computed tomography (CT)-based navigation system for guidance. This study aimed to evaluate the clinical accuracy and precision of HRA performed using the CT-based navigation systems. HRA was performed on 17 hips guided by the CT-based navigation systems. We measured cup alignment deviation, deviation of the stem position, and alignment from the plan by image matching between pre-operative and post-operative CT images. Cup anteversion was within 5° of that in the plan in all cases. Cup inclination was within 5° of that in the plan in 82.4% and within 10° in all cases. The angular difference of the stem was within 5° in all cases, and the entry point of the stem was within 4mm in all cases. The CT-based navigation system for HRA guided accurate component placement according to the plan.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.