Abstract

Alongside the development and testing of new audible alarms intended to support International Electrotechnical Commission 60601-1-8, a global standard concerned with alarm safety, the categories of risk that the standard denotes require further thought and possible updating. In this article, we revisit the origins of the categories covered by the standard. These categories were based on the ways that tissue damage can be caused. We consider these categories from the varied professional perspectives of the authors: human factors, semiotics, clinical practice, and the patient or family (layperson). We conclude that while the categories possess many clinically applicable and defensible features from our range of perspectives, the advances in alarm design now available may allow a more flexible approach. We present a three-tier system with superordinate, basic, and subordinate levels that fit both within the thinking embodied in the current standard and possible new developments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.