Abstract
Implantable cardioverter defibrillators (ICD) are typically programed with multiple treatment zones and discriminators to minimize inappropriate therapies for supraventricular tachycardia while still delivering life saving therapies for ventricular tachycardia (VT) and fibrillation (VF). Biotronik ICDs freeze the VT counters when tachycardia is in the VF zone due to lack of discriminators in the VF zone, which may result in an inappropriate delay in tachycardia detection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.