Iothalamate (IOT) clearance and p-aminohippuric acid (PAH) clearance are used for estimation of glomerular filtration rate (GFR) and effective renal plasma flow (ERPF). A simple and rapid method is described for simultaneous determination of IOT and PAH in the same chromatogram in the serum and urine of humans. The mobile phase consisted of methanol–50 m M sodium monobasic phosphate with 0.5 m M tetrabutyl ammonium hydrogen sulfate (18:82, v/v), pumped at a rate of 0.8 ml/min on a C 1 8 reversed-phase column. Samples of serum and urine were deproteinized with two volumes of acetonitrile containing the internal standard, p-aminobenzoic acid (PABA). The UV detector was set at 254 nm and peak height ratios of PAH or IOT to PABA were calculated with an integrator. Precision and accuracy were within 15% for both PAH and IOT. The recovery of PAH in urine and serum were 94% and 91%, respectively. For IOT the corresponding recoveries were 93% and 92%, respectively. This method clearly distinguishes acetyl-PAH from PAH and has been validated in healthy volunteers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.