Abstract

Superwarfarins are second-generation long-acting anticoagulant rodenticides that can cause unintended human and wildlife toxicity due, in part, to their prolonged half-lives. Commercially available superwarfarin rodenticides are synthesized as racemates with two asymmetric carbons, producing four stereoisomers. To support studies of human plasma half-lives of individual superwarfarin stereoisomers, a method was developed based on LC-MS/MS to separate and quantify stereoisomers of the commercially important superwarfarins bromadiolone, difenacoum and brodifacoum. Human plasma samples were prepared using protein precipitation and centrifugation. Chiral-phase HPLC separation was carried out on-line with tandem mass spectrometric quantitative analysis of the eluting stereoisomers using selected-reaction monitoring with positive ion electrospray on a triple quadrupole mass spectrometer. All four stereoisomers of each superwarfarin were resolved within 12.5 min with calibration curves spanning 2–3 orders of magnitude and lower limits of quantitation between 0.87 and 2.55 ng/mL. This method was used to determine the half-lives of superwarfarin stereoisomers in plasma from patients who had inhaled synthetic cannabinoid products contaminated with superwarfarins. These data may be used to guide the development of safer next generation anticoagulant rodenticides stereoisomers.