Abstract

A coupled system for the online detection of total arsenic is presented consisting of a flow injection system (high pressure liquid (HPL) pump, injection loop of 250μL) and high-resolution continuum source hydride generation atomic absorption spectrometry (HR-CS-HG-AAS).The system enables the continuous detection of volatile arsenic compounds and is adaptable for a wide range of elements.The analytical method was applied for the determination of total arsenic content in human urine. The limit of detection (LOD) with 0.09μg/L (peak height) and the limit of quantification (LOQ) with 0.37μg/L are comparable to standard methods.In this work, possibilities of arsenic determination with FI–HR-CS–HG–AAS have been explored. The online-coupling of a flow injection system with HR-CS–HG–AAS as an element-specific detector is a novel method for the detection and quantitative determination of arsenic and was reported for the first time in this paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.