Abstract

CIEF was developed for the rapid analysis of charge heterogeneity of trastuzumab biosimilar using commercially available fluorocarbon-coated capillary. The CIEF master mix was composed of 0.30% w/v methyl cellulose, 2.3 M urea, 56.8 mM l-arginine, 1.52 mM iminodiacetic acid, 4.5% v/v carrier ampholytes (broad-range pI 3-10 and narrow-range pI 8-10.5 with ratio of 3:1), and 0.45% v/v 10.0, 9.5, 7.0, 5.5, 4.1 pI markers. To get a robust method to analyze charge heterogeneity, some separation parameters, including focusing time and separation temperature, were investigated and optimized. The optimized method gave good precision in estimated pI values of charge variants with RSDs of not more than 0.16% intraday analysis (n = 6) and < 0.18% interday analysis (n = 9). In addition, the applications of this method including purity, stability, lot consistency, peptide N-glycosidase F digest, and C-terminal lysine variants characterization were also investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.