Abstract

Size-exclusion chromatography (SEC) provides a rapid method for determining molar mass distributions of macromolecules. In this article, we demonstrate the method of SEC coupled with an on-line Fourier-transform infrared (FT-IR) detector and an on-line differential viscometer developed for copolymer characterization. Using a chemometric technique to analyze the FT-IR spectra allows for determining copolymer concentration and composition. The combination of the FT-IR concentration detector and the viscometer allows the universal calibration method to be used for measuring copolymer molar mass. Molar mass distribution and compositional distribution are described and the method's application demonstrated using several types of styrene copolymers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.