Abstract

Radius and area function of a spherical indenter are calibrated by performing spherical nanoindentation tests on fused silica with known properties. The spherical indenter is made of ruby with a relatively large radius of 200 μm in nominal value. Hertzian relation was found to be successful in characterizing radius of spherical indenter under purely elastic deformation condition. Oliver and Pharr's method is dependent on the measurement accuracy of residual displacement, which cannot be measured within acceptable precision under very little plastic deformation with values of residual displacement being only several nanometers. Area function calibrated with Oliver and Pharr's method on a brittle material like fused silica was found to be smaller than true value, leading to measured elastic modulus larger than true value. A softer material that can produce sufficient plastic deformation with large residual displacement is recommended as the standard material for area function calibration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.