Abstract
Silicon microstrip detectors are widely used in experiments for space astronomy. Before the detector is assembled, extensive characterization of the silicon microstrip sensors is indispensable and challenging. This work electrically evaluates a series of sensor parameters, including the depletion voltage, bias resistance, metal strip resistance, total leakage current, strip leakage current, coupling capacitance, and interstrip capacitance. Two methods are used to accurately measure the strip leakage current, and the test results match each other well. In measuring the coupling capacitance, we extract the correct value based on a SPICE model and two-port network analysis. In addition, the expression of the measured bias resistance is deduced based on the SPICE model.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Nuclear Science and Techniques
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.