Abstract

The efficiency of implantable electrodes for cardiac pacing depends on the ratio of the conductive surface area to the geometric area of the interface with excitable tissue. New models of heart pacers require reduction of post-pulse polarization, i.e. the potential left on the electrode / tissue interface after a pacemaker pulse. Increasing the conductive surface area is an effective method to this end. Microscopy provides an important tool in elucidating the role of surface structure in electrode performance.Three different surface textures were characterized on a 90% platinum(Pt)/10% iridium (Ir) polished electrode substrate of roughly 5 mm2 geometric surface area. These consisted of the polished substrate itself, a thin film of textured platinum in the 1-3 micron size range, and a sub-micron platinum black coating. Sample surface effects were characterized via scanning electron microscopy (SEM), in-vitro electrical impedance spectrum analysis, and polarization after-potential measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.