Abstract

We report on the delivery of low energy ultra-short (<1 ps) laser pulses for laser induced breakdown spectroscopy (LIBS). Ultra-short pulses have the advantage of high peak irradiance even at very low pulse energies. This opens the possibility to use compact, rare-earth doped fiber lasers in a portable platform for point detection applications using LIBS for elemental analysis. The use of low energy ultra-short pulses minimizes the generation of a broad continuum background in the emission spectrum, which permits the use of non-gated detection schemes using very simple and compact spectrometers. The pulse energies used to produce high-quality LIBS spectra in this investigation are some of the lowest reported and we investigate the threshold pulse requirements for a number of near IR pulse wavelengths (785-1500 nm) and observe that the pulse wavelength has no effects on the threshold for observation of plasma emission or the quality of the emission spectra obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.