Abstract
The response of lithium fluoride (LiF) crystal detectors to monochromatic X-rays is measured in the multi-kilo-electron-volt range. This response, as a function of the X-ray dose, is independent of photon energy with no saturation level found. The response, as a function of the incident energy flux, is found to increase for photons of lower energy due to the differing attenuation lengths of X-ray photons within the crystal. Small differences are seen between different confocal microscopes used to scan the data, suggesting the need for absolute calibration. The spatial resolution of the LiF is also measured (1.19-1.36 μm) and is found to be independent of incident photon energy. Finally, a photometric study is performed in order to assess the feasibility of using these detectors at current X-ray free electron laser and laser facilities worldwide.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.