Abstract

A two-component thermoluminescent albedo neutron monitoring system was developed at Instituto de Radioproteção e Dosimetria, Brazil. As there is no Brazilian regulation for neutron individual monitoring service, the system was tested according to the ISO 21909 standard. This standard provides performance and test requirements for determining the acceptability of personal neutron dosemeters to be used for the measurement of personal dose equivalent, H p (10), in neutron fields with energies ranging from thermal to 20 MeV. Up to 40 dosemeters were used in order to accomplish satisfactorily the requirements of some tests. Despite operational difficulties, this albedo system passed all ISO 21909 performance requirements. The results and problems throughout this characterization are discussed in this paper.