Abstract

The uncertainties and the stability of a neutron sensitive MCP/Timepix detector when operating in the event timing mode for quantitative image analysis at a pulsed neutron source were investigated. The dominant component to the uncertainty arises from the counting statistics. The contribution of the overlap correction to the uncertainty was concluded to be negligible from considerations based on the error propagation even if a pixel occupation probability is more than 50%. We, additionally, have taken into account the multiple counting effect in consideration of the counting statistics. Furthermore, the detection efficiency of this detector system changes under relatively high neutron fluxes due to the ageing effects of current Microchannel Plates. Since this efficiency change is position-dependent, it induces a memory image. The memory effect can be significantly reduced with correction procedures using the rate equations describing the permanent gain degradation and the scrubbing effect on the inner surfaces of the MCP pores.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.