Abstract

Measurements were made of the characteristics of the Gatan PEELS system, to evaluate how this parallel-recording spectrometer can be used to best advantage. An increase in sensitivity at higher signal level is attributable to the electronics and can be largely removed by a minor modification to the circuitry. The signal/noise ratio and DQE can be understood only if explicit allowance is made for the point-spread function of the detector. DQE is found to exceed 0.1 over a limited range of signal intensity. The stray-scattering background, response time of the detector and other spectrometer properties have also been measured.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.