Abstract

Inherent aberrations of gradient index (GRIN) lenses used in fluorescence endomicroscopes deteriorate imaging performance. Using adaptive optics, we characterized and corrected the on-axis and off-axis aberrations of a GRIN lens with NA 0.8 at multiple focal planes. We demonstrated a rotational-transformation-based correction procedure, which enlarged the imaging area with diffraction-limited resolution with only two aberration measurements. 204.8 × 204.8 µm2 images of fluorescent beads and brain slices before and after AO corrections were obtained, with evident improvements in both image sharpness and brightness after AO correction. These results show great promises of applying adaptive optical two-photon fluorescence endomicroscope to three-dimensional (3D) imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.