Abstract

An absolute determination of the absorbed dose rates to tissue from a Sr90Sr+90Y β-ray secondary standard source was attempted using an extrapolation chamber developed at the Calibration Laboratory of São Paulo. The performance of this chamber and its main characteristics are presented in this work. A comparison of the results obtained with those of the calibration certificate indicates an absorbed dose rate agreement within 1.0%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.