Abstract

Abstract Output measurements in terms of absorbed dose rate to tissue were carried out for the ISO (International Organization for Standardization) recommended beta sources 90Sr/90Y, 85Kr and 147Pm using an extrapolation chamber. A holder made of anodized aluminium for the extrapolation chamber for the output measurements was designed indigenously. The measured dose rates for three beta sources were compared with the corresponding values provided in the certificate by the Physikalisch-Technische Bundesanstalt (PTB), the National Standard Laboratory of Germany. Depending upon the source and source-to-chamber distance, the PTB certificate values have uncertainties in the range ±1.2–±1.5%. The measured dose rates have uncertainties of about ±0.3%. The deviations between the measured dose rates and the PTB certificate values compare well (maximum deviation is about 2%) considering the combined uncertainties associated with the measurements and the certificate values. The study suggests that the beta irradiation system comprising the above beta sources can be utilized for quality assurance programme related to personnel monitoring services in India.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.