Abstract

The development of a new boron compound for boron neutron capture therapy (BNCT) requires a high-intensity thermal neutron beam and the construction of thermal neutron fields in place of research reactors scheduled for suspension in the future. A thermal neutron irradiation field was produced by combining a cyclotron-based epithermal neutron source with a moderator to form an irradiation field for mice and cells, and its characteristics were evaluated.Our study confirmed that the lower limbs of the mouse could be sufficiently irradiated with a thermal neutron intensity of 4.83 × 108 n cm−2 s−1. The neutron dose rate and gamma ray dose rate per thermal neutron flux were 8.3 × 10−13 and 1.4 × 10−12 Gy cm2, respectively. In addition, by using a water phantom, irradiation with a maximum thermal neutron flux of 1.35 × 109 n cm−2 s−1 was achieved. In this case, the neutron dose rate and gamma ray dose rate per thermal neutron flux were 2.7 × 10−13 and 8.3 × 10−13 Gy cm2, respectively. An irradiation field capable of irradiating mice and cells was successfully produced and compared favorably with the irradiation fields from research reactors currently in use.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call