Abstract

A pulse-dilation photomultiplier tube (PD-PMT) with sub-20 ps temporal resolution has been developed for use with γ-ray-sensitive gas Cherenkov detectors at the National Ignition Facility to improve the diagnosis of nuclear fusion burn history and the areal density of the remaining capsule ablator. The pulse-dilation mechanism entails the application of a time-dependent, ramp waveform to a photocathode-mesh structure, introducing a time-dependent photoelectron accelerating potential. The electric field imparts axial velocity dispersion to outgoing photoelectrons. The photoelectron pulse is dilated as it transits a drift region prior to amplification in a microchannel plate and read out with a digital oscilloscope. We report the first measurements with the prototype PD-PMT demonstrating nominal <20 ps FWHM across a 400 ps measurement window and <30 ps FWHM for an extracted charge up to 300 pC. The output peak areas are linear to within 20% over 3 orders of magnitude of input intensity. 3D particle in cell simulations, which included space charge effects, have been carried out to investigate the device temporal magnification, resolution, and linearity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.