Abstract
PurposeTo utilize a novel ex vivo perfused human renal model and quantify microwave ablation (MWA) size differences in renal tissue when combining MWA with transarterial embolization (TAE). Materials and MethodsHuman kidneys (n = 5) declined for transplantation were obtained and connected to a fluoroscopy-compatible ex vivo perfusion system. Two ablations—1 standard MWA and 1 TAE-MWA—were performed in each kidney for 2 minutes at 100 W using a MWA system (Solero Angiodynamics). MWA alone was performed in the upper pole. In the lower pole, MWA was performed after TAE with 40–90 μm radiopaque microspheres to achieve angiographic stasis. Ablation zones of coagulative necrosis were sectioned along the long axis and segmented for maximal short-axis diameter (SAD) and long-axis diameter (LAD) measurements. ResultsA total of 10 ablations (5 MWAs and 5 TAE-MWAs) were performed in 5 human kidneys. TAE-MWA resulted in significantly increased SAD, LAD, volume, and sphericity compared with standard MWA ± SD, with mean measurements as follows (5 standard MWAs ± SD vs 5 TAE-MWAs ± SD, 2-tailed t-test): (a) SAD, 1.8 cm (SD ± 0.1) versus 2.5 cm (SD ± 0.1) (P < .001); (b) LAD, 2.9 cm (SD ± 0.3) versus 3.2 cm (SD ± 0.1) (P = .039); (c) volume, 5.0 mL (SD ± 0.5) versus 11.0 mL (SD ± 0.7) (P < .001); and (d) sphericity, 0.4 (SD ± 0.2) versus 0.6 (SD ± 0.1) (P = .049). Histology demonstrated no differences in TAE-MWA other than concentrated microspheres. ConclusionsThis ex vivo human kidney perfusion model confirmed that combined MWA-TAE significantly increased ablation size and spherical shape compared with MWA alone.