Abstract
Only extreme degrees of endotracheal tube (ETT) narrowing can be detected with monitoring of tidal volume (V(T)) during pressure-controlled ventilation (PCV). To assess the degree of ETT obstruction in PCV and to compare it to V(T) monitoring, we produced 3 levels of partial ETT obstruction in 11 healthy anesthetized piglets using ETTs of 4 different inner diameters (IDs 9.0, 8.0, 7.0, and 6.0 mm). An expiratory flow over volume ((e)-V) curve was plotted and the time constant (tau(e)) at 15% of expiration time (T(e)) was calculated. We also calculated the fractional volume expired during the first 15% of T(e) (V(ex fract,15)) and compared those variables to full expiratory V(T) for each of the 3 obstructions. V(T) monitoring failed to detect ETT narrowing. By contrast, V(ex fract,15) decreased and tau(e) increased significantly with increasing ETT narrowing (for IDs 9.0, 8.0, 7.0, and 6.0, mean V(ex fract,15) was 195, 180, 146, and 134 mL respectively and mean tau(e) was 380, 491, 635, 794 ms for IDs 9.0, 8.0, 7.0, and 6.0 respectively). We conclude that when the elastic recoil that drives (e) is appropriately considered, analysis of (e) and V(ex fract,15) detects partial ETT obstruction during PCV.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.