Abstract

Rapid airway occlusions during mechanical ventilation are followed immediately by high-frequency pressure oscillations. To answer the question if the frequency of forced pressure oscillations is an indicator for partial obstruction of the endotracheal tube (ETT) we performed mathematical simulations and studies in a ventilated physical lung model. Model-derived predictions were evaluated in seven healthy volunteers. Partial ETT obstruction was mimicked by decreasing the inner diameter (ID) of the ETT. In the physical model ETTs of different ID were used. In spontaneously breathing volunteers viscous fluid was applied into the ETT's lumen. According to the predictions derived from mathematical simulations, narrowing of the ETT's ID from 9.0 to 7.0mm decreased the frequency of the pressure oscillations by 11% while changes of the respiratory system's compliance had no effect. In volunteers, a similar reduction (10.9%) was found when 5 ml fluid were applied. We conclude that analysis of pressure oscillations after flow interruption offers a tool for non-invasive detection of partial ETT obstruction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.