Abstract

The aim of this study was to evaluate the clinical practice of CR mammography in Tanzania. The equipment performance and operational conditions were studied; and mean glandular dose (DG) estimated to 75 women undergoing diagnosis at three mammography facilities. All mammograms during this study were reported to be useful for the intended diagnosis. The median DG for craniocaudal and mediolateral oblique projections ranged from 1.27 ± 0.18 mGy to 1.9 ± 0.27 mGy and from 1.3 ± 0.18 mGy to 1.9 ± 0.27 mGy, respectively, and were below the national regulatory guidance of 2.5 mGy. Despite this positive result, unavoidable inappropriate use of beam quality and tube loading settings which could have been through appropriate staff training and performing routine quality control were not uncommon. This work provides an insight of current operational conditions of CR in Tanzania and what strategy should be employed to this service to improve patient care in the country.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.