The authors report on 2 cases of delayed compression of the cervical spinal cord by dense scar tissue forming around epidural electrodes implanted for spinal cord stimulation (SCS). This complication has not been previously reported. Myelopathy developed in 2 patients 5 and 16 years after the surgical placement of a cervical epidural electrode. Prior to myelopathy, both patients experienced a tolerance phenomenon. Magnetic resonance imaging in both cases showed severe compression of the spinal cord by the electrode. At surgery, thick scar tissue surrounding the electrode and compressing the posterior aspect of the cord was discovered. Both patients experienced a full recovery following the removal of both scar tissue and the electrode. Medical and paramedical staff dealing with SCS should be aware of this severe and delayed complication. In addition, the development of epidural fibrosis can explain the occurrence of tolerance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.