Abstract
Hydromer's polymeric formulations F200 and F202 were evaluated after application to a synthetic substrate for effects on cell adhesion and proliferation. A significant reduction in cell adhesion was observed when cells grown on medical-grade stainless steel coated with these polymers were stained and examined under a fluorescence microscope. This reduction in cell adhesion/proliferation was confirmed when cells were isolated and analyzed by the MTS cell proliferation assay. The rate of growth of cells on F200- and F202-coated stainless steel monitored over a period of 7 days was significantly less than that observed on uncoated stainless steel, suggesting that the rate of growth of cells was reduced. The adhesion/proliferation of human umbilical vein endothelial cells (HUVEC) to coated substrates was also decreased significantly, indicating that the reduction in cell adhesion/proliferation is not restricted to only fibroblasts. Additional studies have indicated that the adhesion/proliferation of murine fibroblasts and human endothelial cells to stainless coated with a modified formulation exhibiting a high degree of lubricity was also significantly reduced. This lubricious formulation was also observed to be effective in reducing platelet adhesion, data supporting the view that lubricity also contributes to a reduction in cell and platelet adhesion. The application of these polymeric coatings on devices designed for medical implantation may not only prevent thrombus formation but may also retard the process of restenosis.