Abstract

Extracorporeal membrane oxygenation (ECMO) is used in critically ill patients that require respiratory and/or cardiac support. Cefiderocol is a novel siderophore antibiotic that may require use in infected critically ill patients supported by ECMO. The objective of this study was to determine the loss of cefiderocol through an ex vivo adult ECMO circuit using a Quadrox-iD oxygenator. A 3/8-inch, simulated, ex vivo closed-loop ECMO circuit was prepared with a Quadrox-iD adult oxygenator and primed with fresh whole blood. Cefiderocol was administered into the circuit to achieve a starting concentration of approximately 90 mg/L. Post-oxygenator blood samples were collected at 0, 0.25, 0.5, 1, 2, 4, 6, 12, and 24 h after the addition of the drug to determine the loss in the circuit. A glass control jar was prepared with the same blood matrix and maintained at the same temperature to determine drug degradation. The experiment was conducted in triplicate. The rate of cefiderocol loss in the ECMO circuit was compared with that in the control by one-way analysis of variance. At 0h, the difference between the pre- and post-oxygenator concentrations was -4 ± 4% (range 0 to -7%). After 24h, the cefiderocol percent reduction was similar between the ECMO circuit and control (50% ± 13 vs. 50% ± 9, p=1.0). The degradation rate of cefiderocol did not differ significantly within the ECMO circuitand control, suggesting no loss due to sequestration or adsorption. Pharmacokinetic studies in patients supported by ECMO are warranted to determine final dosing recommendations.