Abstract

Charge heterogeneity analysis of monoclonal antibodies (mAbs) and complex formats, such as bispecifics, is crucial for therapeutic applications. In this study, we developed two capillary electrophoresis (CE)-based methods, capillary zone electrophoresis (CZE) and imaged capillary isoelectric focusing (iCIEF), for analyzing a broad spectrum of mAbs and complex mAb formats. For CZE, we introduced a new buffer system and optimized the background electrolyte (BGE) with an alternative dynamic coating agent and a superior polymeric additive. The pH of the BGE was increased, leading to enhanced resolution of high pI and complex format mAbs. In iCIEF, we identified an ampholyte combination offering a highly linear pH gradient and covering a suitable pH range. We also investigated alternatives to denaturing stabilizers and found that non-detergent sulfobetaine 195 exhibited excellent properties for iCIEF applications. These optimized methods provide a framework for the charge heterogeneity analysis of therapeutic mAbs and complex formats.