Abstract

Complications of ceramic-polyethylene bearing total hip arthroplasty (THA) include osteolysis, loosening, dislocation, and component failure. Catastrophic acetabular component failure involves severe damage to both the polyethylene liner and metal shell. This case study presents the first reported complete wear-through of the acetabular portion of a ceramic-polyethylene arthroplasty presenting as a dislocation and a review of the literature. In this study, a patient's alumina ceramic femoral head penetrated the polyethylene liner and titanium shell and presented as a dislocated THA. The contributing factors for this catastrophic failure include young patient age, high activity level, thin polyethylene liner, backside wear, component positioning, polyethylene sterilization with γ irradiation in air, and lack of appropriate follow-up. Revision THA was performed without complications.